The New England Journal of Medicine
e-mail icon  FREE NEJM E-TOC    HOME   |   SUBSCRIBE   |   CURRENT ISSUE   |   PAST ISSUES   |   COLLECTIONS   |    Advanced Search
Sign in | Get NEJM's E-Mail Table of Contents — Free | Subscribe
 
Original Article
Volume 311:485-490 August 23, 1984 Number 8
NextNext

Early experience with an implantable cardioverter
DP Zipes, JJ Heger, WM Miles, Y Mahomed, JW Brown, SR Spielman, and EN Prystowsky

 Sign up for free e-toc
 

Tools and Services
-Add to Personal Archive
-Add to Citation Manager
-Notify a Friend
-E-mail When Cited

More Information
-PubMed Citation
Abstract

We tested the efficacy and safety of a fully programmable cardioverter weighing 95 g, in terminating sustained ventricular tachycardia. The device was implanted tranvenously under local anesthesia in seven patients. On command from a programmer or automatically, the cardioverter delivered shocks through a lead inserted to the apex of the right ventricle. It also served as a demand ventricular pacemaker and could perform programmed ventricular stimulation or overdrive pacing. Cardioversion of ventricular tachycardia required less than 0.5 J (mean) and was well tolerated by the patients, who were awake and not sedated. In one patient, a shock terminated ventricular tachycardia with the device in the automatic mode but produced atrial fibrillation with a rapid ventricular response that was intermittently recognized as ventricular tachycardia, triggering additional shocks. One such shock in the ST segment produced ventricular fibrillation that was terminated transthoracically in the emergency room, without residual impairment. We conclude from these preliminary observations that cardioversion of sustained ventricular tachycardia by means of an implantable catheter device is feasible, but for the present its use in the automatic mode must be cautious and selective. The unit's small size, ease of implantation, usefulness for noninvasive electrophysiologic studies, programmability, and bradycardia pacing functions are advantages. The next-generation device must be able to defibrillate and provide better differentiation of arrhythmias.

This article has been cited by other articles:



HOME  |  SUBSCRIBE  |  SEARCH  |  CURRENT ISSUE  |  PAST ISSUES  |  COLLECTIONS  |  PRIVACY  |  HELP  |  beta.nejm.org

Comments and questions? Please contact us.

The New England Journal of Medicine is owned, published, and copyrighted © 2009 Massachusetts Medical Society. All rights reserved.