Background Several outbreaks and pseudo-outbreaks of Pseudomonas aeruginosa and Serratia marcescens infections associated with bronchoscopy have been reported. We conducted an investigation of P. aeruginosa and S. marcescens isolates related to bronchoscopy at a community hospital.
Methods We reviewed the records of all bronchoscopic procedures at the community hospital from July to October 2001. Environmental samples were obtained. Pulsed-field gel electrophoresis (PFGE) was performed on isolates of P. aeruginosa.
Results From July 1 to October 31, 2001, 66 bronchoscopic procedures were performed in 60 patients, and 43 specimens were obtained for bacterial culture; 20 of the specimens (47 percent) were positive for P. aeruginosa. Six (30 percent) of the specimens that were positive for P. aeruginosa also yielded S. marcescens. All 20 P. aeruginosa isolates were associated with procedures performed with three of four new bronchoscopes from the same manufacturer. Contrary to manufacturing specifications, the biopsy-port caps on all four bronchoscopes were easily removable, and P. aeruginosa was cultured from the biopsy ports of the three implicated bronchoscopes. The PFGE patterns of P. aeruginosa isolates from the bronchoscopes, patients, and two environmental samples were indistinguishable. One patient was hospitalized with P. aeruginosa pneumonia 11 days after bronchoscopy. The manufacturer reported a design change instituted in 1997, and production problems may have resulted in the distribution of bronchoscopes that did not meet specifications.
Conclusions We documented contamination of bronchoscopes with P. aeruginosa and S. marcescens and possible infection of patients at a community hospital as a result of the inadequate disinfection of bronchoscopes because of a manufacturing defect.
In September 2001, an infection-control practitioner at a community hospital notified the Tennessee Department of Health of an increase in the number of P. aeruginosa and S. marcescens isolates associated with bronchoscopy. We report the results of the subsequent investigation.
Methods
We reviewed laboratory reports of all P. aeruginosa and S. marcescens isolates at the community hospital during 2001. A case was defined by the isolation of P. aeruginosa or S. marcescens from a specimen obtained during bronchoscopy in a patient at the community hospital from July 1 to October 31, 2001. We reviewed the endoscopy records, which included the serial numbers of the bronchoscopes used, for all patients who underwent bronchoscopy during this period. We reviewed the medical records of all patients who underwent bronchoscopy during the study period to determine whether any were readmitted for infections caused by P. aeruginosa or S. marcescens during the month after the procedure.
Environmental samples were obtained from the bronchoscopes and other sites within the endoscopy suite. Specimens were obtained from the bronchoscopes by flushing them with sterile saline or by using moistened swabs. Swab specimens were obtained from the automated endoscope reprocessor and items used during procedures. In addition, 100-ml specimens of liquids were collected, including source water, rinse water, and disinfectant used in the automated endoscope reprocessor and tap water and drain water from the cleaning sink. The hospital's endoscope-reprocessing procedures were reviewed, and the bronchoscopes were inspected.
Isolates of P. aeruginosa were compared with the use of pulsed-field gel electrophoresis (PFGE). Restriction-endonuclease digestion of bacterial chromosomes was performed with XbaI (New England Biolabs), and the relatedness of strains was examined with the use of Molecular Analyst software.19
We performed bivariate statistical analyses using chi-square tests calculated with Epi Info 2000 software.20
Results
The community hospital is a 203-bed facility with four pulmonologists who performed 267 bronchoscopy procedures in 2001. From July 1 to October 31, 2001, 66 bronchoscopic procedures were performed in 60 patients with eight different bronchoscopes owned by the hospital, and 43 specimens were obtained for bacterial culture. Of the 43 bacterial cultures, 20 specimens (47 percent) yielded P. aeruginosa. Six of the 20 specimens (30 percent) with P. aeruginosa also yielded S. marcescens.
Of the 20 patients with specimens positive for P. aeruginosa, 11 (55 percent) were male; the median age of these patients was 59 years (range, 24 to 88). Evaluation of pneumonia was the primary indication for bronchoscopy in the case of only four patients (20 percent). None of the 20 patients were discharged with a diagnosis of pneumonia caused by P. aeruginosa or S. marcescens.
Between July 1 and September 17, when the investigation began, four of the hospital's eight bronchoscopes had been used, but P. aeruginosa was isolated from specimens obtained from only two (bronchoscopes A and B). Both were new bronchoscopes of similar models (BF-1T160 and BF-160, respectively) produced by the same manufacturer (Olympus America).
Bronchoscopes A and B had been received by the hospital on July 11, just eight days before the collection of the first specimen that was positive for P. aeruginosa (Figure 1). They were first used on July 13 and July 24, respectively, and specimens positive for P. aeruginosa were obtained by the time of the third procedure performed with either bronchoscope. Of the 16 specimens collected for bacterial culture from patients during bronchoscopy with bronchoscope A, 12 (75 percent) were positive for P. aeruginosa; 5 of 8 specimens (62 percent) collected from patients during bronchoscopy with bronchoscope B were positive for P. aeruginosa. Bronchoscopes A and B were removed from service on September 18, and no P. aeruginosa isolates were reported in specimens obtained during bronchoscopy during the following five weeks.
|
|
|
The four recent-model bronchoscopes with loose biopsy-port caps (bronchoscopes A, B, C, and D) were used for 40 procedures; 28 specimens were obtained for culture, and 20 were positive for P. aeruginosa (Table 1). We compared these bronchoscopes with four older-model bronchoscopes by the same manufacturer (bronchoscopes E, F, G, and H) that were in use during the study period and that were never noted to have loose or removable biopsy-port caps (Table 1). The risk that a specimen would be positive for P. aeruginosa was associated only with the use of recent-model bronchoscopes with the loose biopsy-port caps (P<0.001).
|
|
The bronchoscopes were routinely processed in an automated endoscope reprocessor (model MV-2, MediVators) with use of a 0.55 percent solution of ortho-phthalaldehyde (Cidex OPA, Advanced Sterilization Products). A representative of the manufacturer of the automated endoscope reprocessor reviewed the hospital's reprocessing procedures and indicated that they were satisfactory, except that the source water filter for the reprocessor had not been changed within the recommended six months. No strains of P. aeruginosa were isolated from either the source water or the filter.
A 50-year-old woman was readmitted 11 days after bronchoscopy with new pulmonary infiltrates in the right middle lobe. The specimen obtained during bronchoscopy was positive for P. aeruginosa. On readmission, she had fever, leukocytosis, dyspnea, and increasing cough with darkening sputum. Culture of a sputum sample yielded P. aeruginosa with an antibiotic-susceptibility pattern similar to that of the epidemic strain. No isolates were available for PFGE. The pneumonia resolved.
Discussion
A striking increase in the number of isolates of P. aeruginosa and S. marcescens from bronchoscopy specimens occurred at a community hospital from July through October 2001. One bronchoscopy-associated case of P. aeruginosa pneumonia was identified. The likelihood of positivity for P. aeruginosa or S. marcescens was associated only with the use of recent-model bronchoscopes that had loose biopsy-port caps. Strains of P. aeruginosa and S. marcescens were isolated from the biopsy ports of the implicated bronchoscopes. PFGE patterns linked the P. aeruginosa isolates from the patients to the isolates from the bronchoscopes.
Because the risk of a specimen's being positive for P. aeruginosa or S. marcescens was associated exclusively with the use of the recent-model bronchoscopes with the loose biopsy-port caps and because a case of bronchoscopy-associated pneumonia was identified, the manufacturer and the Food and Drug Administration were notified on September 18 and October 8, respectively. At that time, the Centers for Disease Control and Prevention indicated that no related outbreaks had been reported.
In 1997, the manufacturer changed the design of the biopsy port on at least 15 models of bronchoscopes that were still in use in 2001. Part of the new design involved changing the biopsy-port housing to a removable cap that was screwed on but fixed with adhesive. The manufacturer reported that failure to conform to production standards resulted in the distribution of bronchoscopes that did not meet manufacturing specifications. This might explain why problems with these models of bronchoscopes had not been reported previously. An alternative explanation is that infectious complications related to bronchoscopes often go unrecognized or unreported.9,21,22,23
Persistent bacterial contamination of the biopsy port was probably not a result of a breach of the cleaning and disinfection guidelines, because the hospital's reprocessing procedures were determined to be satisfactory. In addition, contamination persisted even after three cycles of cleaning and disinfection in the automated reprocessor, and other bronchoscopes in use during the study period were not associated with positive cultures.
The loose biopsy-port caps most likely interfered with three important steps in reprocessing performed according to the current guidelines: mechanical cleaning, high-level disinfection, and drying.7 Mechanical cleaning is necessary to remove the biofilm that can interfere with disinfection.24,25 Contaminated surfaces inside the hollow biopsy-port caps could not be cleaned mechanically. High-level disinfection requires an approved disinfectant to be in contact with a contaminated surface for a specified period of time.26 The loose biopsy-port caps, which probably had air pockets when submerged, might have prevented effective contact between the disinfectant and the microorganisms. Finally, drying is recommended to prevent the growth of microorganisms with a predilection for moist environments.27 The design of the caps probably prevented thorough drying of surfaces inside the caps.
The origin of the contaminating bacteria remains elusive. One postulated sequence of events is that the loose biopsy-port caps were contaminated in the sink when the bronchoscopes were cleaned before being placed in the automated endoscope reprocessor. Once within the biopsy-port caps, the bacteria were protected from the chemical-disinfection process. During subsequent use of the bronchoscopes, sterile saline was flushed through the biopsy ports, thereby becoming contaminated and further contaminating the bronchoscope channels and the patients' lower respiratory tracts. The contaminated lavage specimens were then suctioned back through the bronchoscope channels by means of the suction ports and collected for bacterial culture, yielding false positive results.
Antibiotic therapy was initiated in five patients (25 percent) after false positive culture results were reported. If these results reflected contaminated bronchoscopes rather than actual infections, the patients may have received unnecessary antibiotic treatment. Conversely, if patients were infected during the bronchoscopic procedures, the antibiotics may have prevented nosocomial illnesses.
After receiving our report, the manufacturer of the bronchoscopes initiated a recall of certain models. The voluntary recall, issued on November 30, 2001, affected 15 models of bronchoscopes, including the 2 models implicated in this investigation. The stated purpose was "to address a potential looseness of the biopsy-channel port housing [cap]." The recall reportedly involved an estimated 4700 bronchoscopes nationwide and was later expanded to include approximately 14,000 bronchoscopes worldwide. The manufacturer proposed modifying the bronchoscopes to resolve the problem with the caps. The recall received national media attention after an outbreak associated with the recalled models of bronchoscopes was reported at a hospital in another state, as described by Srinivasan et al.28 Further microbiologic and epidemiologic surveillance is warranted to ensure that bronchoscopes modified as a result of the recall are not associated with further cases of false positive clinical specimens or infections.
We are indebted to Henrietta Hardin and the Tennessee Department of Health State Laboratory staff; to Patricia Coleman, Lynn King, and the community hospital pulmonologists and endoscopy staff; to Nita Lamb and the Vanderbilt University Medical Center infection-control practitioners for consultation and advice; to Walter E. Phillips, Ph.D.; and to Laura J. Fehrs, M.D.
Source Information
From the Epidemic Intelligence Service, Centers for Disease Control and Prevention, Atlanta (D.L.K.); the Tennessee Department of Health, Nashville (D.L.K., T.F.J., A.S.C.); the Vanderbilt University School of Medicine, Nashville (T.F.J., A.S.C., P.S.C., W.S.); and Skyline Medical Center, Nashville (G.G.M., J.A.P.).
Address reprint requests to Dr. Schaffner at the Vanderbilt University School of Medicine, Medical Center North, Rm. A-1124, Nashville, TN 37232, or at william.schaffner{at}vanderbilt.edu.
References
| |||||||||||||||||||||||||||||||||||||||||||||||||||
Related Letters:
Outbreaks of Infection Associated with Bronchoscopes