Background Computer-aided detection identifies suspicious findings on mammograms to assist radiologists. Since the Food and Drug Administration approved the technology in 1998, it has been disseminated into practice, but its effect on the accuracy of interpretation is unclear.
Methods We determined the association between the use of computer-aided detection at mammography facilities and the performance of screening mammography from 1998 through 2002 at 43 facilities in three states. We had complete data for 222,135 women (a total of 429,345 mammograms), including 2351 women who received a diagnosis of breast cancer within 1 year after screening. We calculated the specificity, sensitivity, and positive predictive value of screening mammography with and without computer-aided detection, as well as the rates of biopsy and breast-cancer detection and the overall accuracy, measured as the area under the receiver-operating-characteristic (ROC) curve.
Results Seven facilities (16%) implemented computer-aided detection during the study period. Diagnostic specificity decreased from 90.2% before implementation to 87.2% after implementation (P<0.001), the positive predictive value decreased from 4.1% to 3.2% (P=0.01), and the rate of biopsy increased by 19.7% (P<0.001). The increase in sensitivity from 80.4% before implementation of computer-aided detection to 84.0% after implementation was not significant (P=0.32). The change in the cancer-detection rate (including invasive breast cancers and ductal carcinomas in situ) was not significant (4.15 cases per 1000 screening mammograms before implementation and 4.20 cases after implementation, P=0.90). Analyses of data from all 43 facilities showed that the use of computer-aided detection was associated with significantly lower overall accuracy than was nonuse (area under the ROC curve, 0.871 vs. 0.919; P=0.005).
Conclusions The use of computer-aided detection is associated with reduced accuracy of interpretation of screening mammograms. The increased rate of biopsy with the use of computer-aided detection is not clearly associated with improved detection of invasive breast cancer.
Source Information
From the University of California, Davis, Sacramento (J.J.F.); the National Cancer Institute, Bethesda, MD (S.H.T.); Oregon Health and Science University, Portland (P.A.C.); Group Health Cooperative, Seattle (L.A.); the University of California, San Francisco, San Francisco (E.A.S.); the Emory Clinic, Atlanta (C.D.); Northwestern University, Chicago (E.A.B., R.E.H.); the University of Alabama at Birmingham, Birmingham (G.C.); Cancer Research and Biostatistics, Seattle (W.E.B.); and the University of Washington, Seattle (J.G.E.).
Address reprint requests to Dr. Fenton at the Department of Family and Community Medicine, UC Davis Health System, 4860 Y St., Ste. 2300, Sacramento, CA 95817, or at joshua.fenton{at}ucdmc.ucdavis.edu.
This article has been cited by other articles:
HOME | SUBSCRIBE | SEARCH | CURRENT ISSUE | PAST ISSUES | COLLECTIONS | PRIVACY | HELP | beta.nejm.org Comments and questions? Please contact us. The New England Journal of Medicine is owned, published, and copyrighted © 2008 Massachusetts Medical Society. All rights reserved. |